Class 2 Biological Safety Cabinets

CLASS 1 BIOLOGICAL SAFETY CABINETS

Class 1 Biological Safety Cabinets protect the operator and environment by Effectively preventing the release of airborne particulate contaminants.

BioMAT 1

  • BS EN 12469:2000
  • High Efficiency Filtration
  • UKHSA Type Tested
  • ISO Class 5 Clean Environment

CLASS 1 AIRFLOW

Inward airflow travels through the open front, over the worksurface and is HEPA-filtered before being discharged either by releasing back into the room (recirculating) or through a building extract system.

Class 1 cabinets are suitable for work with all biological agents up to ACDP Hazard Group HG3 and Biosafety Level BSL-3. For level 4 (HG4/BSL-4) or applications requiring fully sealed containment, explore our Isolators or Class 3 Biological Safety Cabinets.

This type of cabinet is not designed to offer protection to material being handled. For product protection, explore our Class 2 Biological Safety Cabinets.

  • Blue: Inward environment air 
  • Green: Filtered clean air 

FEATURES INCLUDE:

  • Pre filters to extend filter life

  • High-efficiency HEPA filtration
  • Energy-efficient LED lighting
  • Real-time cabinet status monitoring
  • Touchscreen display
  • High grade 316 stainless steel worksurface
  • Integrated service connections
  • Energy-saving EC fans
  • Fumigation mode and connections
  • Integral formalin vaporizer
  • UV Disinfection with auto mode
  • Closure panel included
  • Split or single piece worktrays

  • Additional carbon filtration

  • Additional fumigation door

  • Uninterrupted power supply

  • Fixed, mobile, or motorised stand

DESIGNED FOR YOUR PROCESS | BUILT FOR YOUR SUCCESS

Have a feature in mind that’s not shown here?

Class 1 Biological Safety Cabinet for Pathology
Custom Class 1 Biological Safety Cabinet for Pathology Work

Need a custom solution designed specifically for your application?

Electrical Supply
Construction Materials
Service Connections
Configuration

EXHAUST AIRFLOW OPTIONS

  • Ducted exhaust module incorporating single HEPA filter and integral anti blowback damper. 
  • Thimble exhaust module incorporating double HEPA filters and integral anti blowback damper. 

  • Recirculating exhaust module incorporating double HEPA filters and vertical or horizontal discharge grille for return air back into the laboratory. 

  • TriMAT constant exhaust bypass module to simplify the extraction of air from laboratories, exclusive to CAS.

  • Exhaust module can be configured in multiple positions to accommodate your laboratory layout or ducting requirements.

PRODUCT SPECIFICATION

*Typical specification based on standard models.* Product-specific technical data to be advised during the consultation/design stage

Cabinet size  1200mm 1500mm 1800mm
External w/d/h (mm)  1200/630/1250  1500/630/1350  1800/630/1350 
Internal w/d/h (mm)  1195/625/700  1495/625/700  1795/625/700 
Weight  200kg  250kg  300kg
Exhaust air volume (m3/s)  0.140  0.180  0.210 
Re-circ  air volume (m3/s)   0.140 0.180  0.210
Noise level  <60dBA  <62dBA  <65dBA 
Light level  900 lux  900 lux  900 lux 
Duct connection  160mm  200mm  200mm 
Power consumption  0.4kw  0.5kw  0.6kw
Standard  Type tested and approved to EN 12469:2000. Sited to BS 5726:2005. 
Voltage AC  230V 
Filtration  H-14 HEPA (standard) @ 0.3µ 99.995% efficiency (ULPA available) 
Construction  2mm mild steel  
Finish  Traffic white RAL 9016 
Inner Worksurface  Grade 304 stainless steel 
Side Windows 5mm UV resistant toughened glass 
Front visor   5mm toughened glass with 200mm open aperture – closed for fumigation using a closure panel (supplied as standard) or fumigation panel (optional extra) 

DISCOVER YOUR SOLUTION

Speak to one of our experts today

WE ARE TRUSTED BY

WHY CHOOSE CAS?

Receive the solution you need to succeed with our free design service, product training, comprehensive warranty, and lifetime support.

Whether it’s a standard in-stock model or a fully custom solution, our expert team is with you every step of the way.

WHAT OUR CUSTOMERS SAY

I do have to say that the quality of the cabinets is really good, UKHSA are very pleased.

UKHSA

We are very satisfied with our equipment and overall the customer service has been fantastic, we will definitely come back to CAS in future if we need anything.

Southampton Solent University

I would like to express my heartfelt thanks and gratitude for the help and assistance provided to the KEMH engineering team.

KEMH

I wanted to make sure it was recognised that the lads worked their behinds off to make sure I got my cabinets in place before I left for the day. Deliveries like that make my life much easier!!

Newcastle University

I do have to say that the quality of the cabinets is really good, UKHSA are very pleased.

UKHSA

Our service has just concluded, please pass on our thanks to the visiting engineers – we were impressed with their efficiency and how smoothly this service round went.

University of Cambridge

The communications generally from CAS were excellent and Brian was great when on site.

Expervent

The team were extremely hard working with no task being too much trouble, even offering to help other contractors that were working within the small space that we had available to us. They really couldn’t have been any more accommodating.

GSK

Just a quick email to say how grateful we are for how everything ran on Friday, we were very impressed by the speed with which everything was smoothed out and taken care of.

InstilBio

All good service as usual from CAS. Your engineer was very helpful, thorough and professional.

Anonymous