CLASS 1-HYBRID
BIOLOGICAL SAFETY CABINETS
Class 1 Hybrid Biological Safety Cabinets offer flexibility with two configurations; Class 1, and full-barrier containment with an added gloveport closure panel. Both configurations provide optimal protection for operators and the environment when handling potentially hazardous materials, while preventing the release of airborne particulate contaminants.
In the Class 1 configuration, a controlled airflow system draws air through the open front, across the work surface, and HEPA-filters exhaust air.
The full-barrier configuration draws air through the integrated side HEPA-filter, across the work surface, and HEPA-filters exhaust air whilst maintaining negative pressure, providing enhanced protection for occasional high-hazard tasks.
Class 1-Hybrid Biological Safety Cabinets from Contained Air Solutions meet BS EN 12469:2000 standards and can be tailored to your specific requirements.
CLASS 1 AIRFLOW
Inward air travels through the open front, over the worksurface, and is HEPA-filtered before being discharged either by releasing it back into the room (recirculating) or through a building extract system.
FULL BARRIER AIRFLOW
Inward air passes through a HEPA-filter before entering the work area, where it circulates clean air and is HEPA-filtered again before discharge. The discharged air can either be released back into the room (recirculating) or through a building extract system.
Blue: Inward environment air | Green: Filtered clean air
WHY CHOOSE CAS?
Receive the solution you need to succeed with our free design service, product training, comprehensive warranty, and lifetime support.
Whether it’s a standard in-stock model or a fully custom solution, our expert team is with you every step of the way.
FEATURES INCLUDE:
DESIGNED FOR YOUR PROCESS | BUILT FOR YOUR SUCCESS



EXHAUST AIRFLOW OPTIONS
PRODUCT SPECIFICATION
*Typical specification based on standard models* Materials and configurations are fully customisable
| Cabinet size | 1200mm | 1500mm | 1800mm |
|---|---|---|---|
| External w/d/h (mm) | 1200/630/1250 | 1500/630/1350 | 1800/630/1350 |
| Internal w/d/h (mm) | 1195/625/700 | 1495/625/700 | 1795/625/700 |
| Weight | 225kg | 275kg | 325kg |
| Exhaust air volume (m3/s) | 0.200 | 0.260 | 0.320 |
| Noise level | <58dBA | <60dBA | <62dBA |
| Light level | 900 lux | 900 lux | 900 lux |
| Duct connection | 200mm | 250mm | 250mm |
| Power consumption | 0.4kw | 0.5kw | 0.6kw |
| Standard | Compliant with EN 12469:2000. Sited to BS 5726:2005. |
|---|---|
| Voltage AC | 230V |
| Filtration | H-14 HEPA (standard) @ 0.3µ 99.995% efficiency (ULPA available) |
| Construction | 2mm mild steel |
| Finish | Traffic white RAL 9016 |
| Inner worksurface | Grade 304 stainless steel |
| Side windows | 5mm UV resistant toughened glass |
| Front visor | 5mm toughened glass with 200mm open aperture – closed for fumigation using a closure panel (supplied as standard) or fumigation panel (optional extra) |
WE ARE TRUSTED BY
WHAT OUR CUSTOMERS SAY

I do have to say that the quality of the cabinets is really good, UKHSA are very pleased.

We are very satisfied with our equipment and overall the customer service has been fantastic, we will definitely come back to CAS in future if we need anything.

I would like to express my heartfelt thanks and gratitude for the help and assistance provided to the KEMH engineering team.

I wanted to make sure it was recognised that the lads worked their behinds off to make sure I got my cabinets in place before I left for the day. Deliveries like that make my life much easier!!

I do have to say that the quality of the cabinets is really good, UKHSA are very pleased.

Our service has just concluded, please pass on our thanks to the visiting engineers – we were impressed with their efficiency and how smoothly this service round went.

The communications generally from CAS were excellent and Brian was great when on site.

The team were extremely hard working with no task being too much trouble, even offering to help other contractors that were working within the small space that we had available to us. They really couldn’t have been any more accommodating.

Just a quick email to say how grateful we are for how everything ran on Friday, we were very impressed by the speed with which everything was smoothed out and taken care of.

All good service as usual from CAS. Your engineer was very helpful, thorough and professional.